规格参数
- 新旧状况
- 新
- Dilution range
- From 1/100 to 1/1000 (50 to 500 ppb with a 50 ppm cylinder)
- Dilution principle
- MFC
- Dilution gas
- Mixtures of SO2 in Air – NO in N2 – CO in Air – HC or BTX in Air Mixtures of several compatible pollutants in the same cylinder
- Diluted gas flow
- 0 to 35 Nml/min air
- Dilution gas line flow accuracy
- +/- 1.5% relative
- Repeatability on flow
- Better than +/- 2,5ppb (at 250ppb O3)
- Stabilization time
- < 5 min
- Internal 0 air quality
- SO2 & O3 < 1 ppb – NO & NO2< 2 ppb – HCnm< 50 ppb HCm< 50 ppb – CO< 50 ppb – BTEX< 1 ppb
- Remaining humidity level
- H2O< 100 ppm (-40°C DP)
- Dimensions
- 19″ 3HE/84 TE , 500mm deep portable casing in option
- Net weight
- 20 Kg
- Carrier gas
- High grade zero air from internal zero air generator
- Flow
- About 3000 Nml/min, other values upon request
- Pressure
- 3 +/- 0.3 bar (40 +/- 4 psig)
- Fittings
- 1/4’’ and 1/8’’ Swagelok
- O3 concentration
- 25 – 500ppb, adjustable by software
- Linearity and accuracy of the mfc
- 0.5 % (rel.) + 0,1 FS
- Repeatability of the mfc
- 0.2 % (rel.) + 0,1 FS
- Rs232
- Standard
- Power supply voltage
- 230V/50-60Hz 115V/60Hz 100V/50-60Hz
- Installed power (max)
- 200W
- 产品编号
- 59850339
描述
The multigas Sonimix 6000 C2 calibrator is designed to calibrate manually or by remote control up to 6 monitors installed in an air pollution monitoring station, for parameters such as SO2, HC, BTX, CO, NOx and O3.
The Sonimix 6000 C2 includes Mass Flow Controllers, GPT, ozone and self re-generable zero air generator and is equipped with
– an inlet for high concentration mixtures of diluted gases such as NO, NO2, SO2, CO, HC, BTX or mixtures of several compatible pollutants in the same cylinder.
– a zero air generator composed of an internal compressor with self re-generable filters, removes following components: moisture, O3, NOx, SO2, HC, BTX and optionally CO and CH4.
– a binary gas dilutor of two inlets (diluted gas and air), including 2 Mass Flow Controllers to generate the zero and other dilutions ratios going from 1/100 to 1/1000 (other values upon request).
The input pressures are monitored by two pressure switches.